316 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Offsets/Limits
Displays menu keys that enable you to configure the offsets and limits for the MCP
measurement.
Key Path: Meas Setup
Remote Command:
There is no equivalent remote command.
Offset A, B, or C
Enables you to select the offset the menu keys will affect.
Key Path: Meas Setup, Offset/Limits
Factory Preset: A
Remote Command: There is no equivalent remote command.
Offset Freq
Sets the frequency difference between the center of the uppermost carrier and the center of
the upper element of the offset pair, as well as the frequency difference between the center
of the lower element of the offset pair and the center of the lower most carrier. If you set the
frequency of any offset to 0 Hz, the offset is turned off and not included in the displayed
results.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: 5 MHz, 10 MHz, 15 MHz
Te rm in a tor s: Hz, kHz, MHz, GHz
Default Terminator: Hz
Range: 0 MHz to 45 MHz
Remote Command:
[:SENSe]:MCPower:OFFSet:LIST:[FREQuency]<Hz>,<Hz>,<Hz>
[:SENSe]:MCPower:OFFSet:LIST:[FREQuency]?