Using the STATus System 27
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 559
Remarks: See also commands :MMEMory:LOAD:STATe and
:MMEMory:STORe:STATe
Example:
*SAV 12
Front Panel Access: File, Save, State
Service Request Enable
*SRE <integer>
*SRE?
Sets the value of the service request enable register.
The query returns the value of the register.
Key Type: There is no equivalent front-panel key.
Range: Integer, 0 to 255
Example: *SRE 22
Read Status Byte Query
*STB?
Returns the value of the status byte register without erasing its contents.
Key Type: There is no equivalent front-panel key.
Remarks: See *CLS
Trigger
*TRG
Triggers the instrument. Use the :TRIGger[:SEQuence]:SOURce command to
select the trigger source.
If you have selected a one-button measurement and it has been paused
(INITiate:PAUSe), or the CONFigure:<meas> command was used, this
command causes the system to exit this “waiting” state and go to the “initiated” state. The
trigger system is initiated and completes one full trigger cycle. It returns to the “waiting”
state on completion of the trigger cycle. See the “Measure (Spectrum Analysis Mode)" on
page 233
for more information about controlling the measurement process.