192 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
9 Input/Output
Factory Preset: Off
Remote Command:
[:SENSe]:SIDentify[:STATe] OFF|ON|0|1
[:SENSe]:SIDentify:[STATe]?
Example:
SID 1
Signal ID Mode
Displays a menu to select the method of signal identification.
Key Path: Input/Output, External Mixer
State Saved: Saved in Instrument State
Factory Preset: Image Suppress
Remote Command:
[:SENSe]:SIDentify:MODE ISUPpress|ISHift
[:SENSe]:SIDentify:MODE?
Example:
SID:MODE ISH
Image Suppress
This signal identification method attempts to suppress all but valid responses by
mathematically removing all image and multiple responses of signals present at the mixer
input. The N8201A internally acquires the data in a two sweep sequence, operates on the
acquired data, and displays the result in Trace 1. Since two measurements are taken for
each display cycle, the display update rate is reduced.
Key Path: Input/Output, External Mixer, Signal ID Mode
State Saved: Saved in Instrument State
Remote Command: See “Signal ID Mode" on page 192.
Example: SID:MODE ISUP