Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 291
Offset/Limits
Accesses menu keys that allow you to configure the offsets for the ACP measurement.
Key Path: Meas Setup
Offset A, B, C, D, E, or F
Accesses a menu of keys to setup an offset.
Key Path: Meas Setup, Offset Setup
State Saved: Saved in Instrument State.
Factory Preset: A
Remote Command: There is no remote command for this function.
Offset Freq
Enables you to set the frequency difference from the center of the main channel to the
center of the offset for a maximum of 6 offsets (labeled A-F). It also allows you to turn
on/off the offsets that you want to measure.
This key label changes to Edge Freq when IS-95C, cdma2000 MC-1X, cdma2000 MC-3X, and
cdma2000 DS is selected.
Key Path: Meas Setup, Offset Setup
State Saved: Saved in Instrument State.
Factory Preset: The following table shows the default array if no radio standard (None)
is selected.
Default Terminator: Hz
Offset State Frequency Integ. BW
A ON 3 MHz 2 MHz
B OFF 0 Hz 2 MHz
C OFF 0 Hz 2 MHz
D OFF 0 Hz 2 MHz
E OFF 0 Hz 2 MHz
F OFF 0 Hz 2 MHz