322 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Power Ref
Enables you to set the multi-carrier power reference to automatic or manual. When set to
automatic, the carrier power result reflects the measured power value in the selected
reference carrier (Meas Setup, Carrier Setup, Ref Carrier). When set to manual, the result is
referenced to the last measured value, or you may specify the reference for the multi-carrier
power measurement. Relative values are displayed, referenced to the “Power Reference”
value.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: Off/Measured power in the reference carrier.
Te rm in a tor s: dBm
Default Terminator: dBm
Range: –200 dBm to 200 dBm
Remote Command:
[:SENSe]:MCPower:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:MCPower:CARRier:AUTO[:STATe]?
[:SENSe]:MCPower:CARRier[:POWer]<dBm>
[:SENSe]:MCPower:CARRier[:POWer]?
Example:
MCP:CARR:AUTO 0
MCP:CARR:AUTO?
MCP:CARR -100
MCP:CARR?