Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 437
Search Param
Displays the search parameter criteria menu that enables you to adjust the parameters for
the peak search functions. These parameters mean that only peaks that rise above the peak
threshold by at least the peak excursion, and then drop by at least the peak excursion, are
identified as peaks.
Key Path: Peak Search
Remote Command:
There is no remote command for this key.
Peak Excursn
Sets the minimum amplitude variation of signals that the marker can identify as a separate
peak. For example, if a peak excursion value of 10 dB is selected, the marker Next Peak
function moves only to peaks that rise more than 10 dB above the Peak Threshold and then
fall back down by at least the Peak Excursn. This criteria applies to all traces. This function
applies to Next Peak, Next Peak Left, and Next Peak Right. If Peak Search is set to Param, it
also applies to Peak Search.
Key Path: Peak Search, Search Param
Dependencies/Couplings: This function is not available when Y-Axis is set to
Frequency instead of Amplitude.
State Saved: Saved in Instrument State.
Factory Preset: 6.0 dB
Te rm in a tor s: dB
Resolution/Rounding/Truncation: 0.01 dB
Range: 0.0 dB to 100 dB
Remote Command:
:CALCulate:MARKer[1]|2|3|4:PEAK:EXCursion <rel_amplitude>
:CALCulate:MARKer[1]|2|3|4:PEAK:EXCursion?
Remote Command Notes:
CALC:MARK:PEAK:SEAR:MODE must be set to PARameter if you want Peak Search
to take advantage of the threshold excursion.