Mode Setup (Spectrum Analysis Mode) 17
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 413
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 412.
Example:
RAD:STAN IS95
J-STD-008
Sets the specific parameters for the selected measurement (located under the “Measure
(Spectrum Analysis Mode)" on page 233 key description) appropriate for industry standard
J-STD-008. The Spectrum Emissions Mask, Harmonic Distortion, MCP, and Spurious
Emissions measurements are not available when this radio standard is selected.
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 412.
Example:
RAD:STAN JSTD
IS-95C
Sets the specific parameters for the selected measurement (located under the “Measure
(Spectrum Analysis Mode)" on page 233 key description) appropriate for industry standard
IS-95C. The IS-95C radio standard is only available when selecting the Channel Power,
Occupied Bandwidth, Adjacent Channel Power, and CCDF one-button measurements.
Key Type: 1 of N menu
Key Path: Mode Setup, Radio Std
Remote Command:
Use[:SENSe]:RADio:STANdard[:SELect] See “Radio Std” on page 412.
Example:
RAD:STAN IS95C