Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 53
Presel Adjust
Allows you to manually adjust the preselector filter center frequency to optimize its
response on the signal of interest. This function is not available in frequency band 0. It is for
signals 2.85 GHz in band 1 and higher. It is not available if the preselector is turned off. The
key also accesses the menu to choose the desired preselector.
For general purpose signal analysis, using Presel Center is recommended. Centering the
filter minimizes the impact of long-term preselector drift. Presel Adjust can be used instead
to manually optimize the preselector. One application of manual optimization would be to
peak the preselector response, which both optimizes the signal-to-noise ratio and
minimizes amplitude variations due to small (short-term) preselector drifting.
The adjustment is preselector specific and the N8201A stores a value for each preselector.
The key readout shows which preselector will be adjusted when the key is pressed. The
choices are [3-26 GHz] and [External]. After performing a Presel Center, the value that is
shown on the Presel Adjust key is the offset from nominal of the affected preselector’s
frequency.
With Presel Adjust selected, press the key again to access the preselector selection menu
to select the preselector you wish to adjust. Some instrument settings will force a
preselector selection. The following flowchart shows the conditions that cause the
preselector selection to change: