432 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
18 Peak Search
Next Pk Left
Moves the marker to the next peak to the left of the current marker. The peak must meet the
defined peak excursion and threshold limits. Peaks that are less than 1% of the current span
away from 0 Hz are ignored. If no valid peak is found, an error “No Peak Found” is
displayed. Press
ESC to clear this message before attempting another search. (Also see
“Peak Excursn" on page 437 and “Pk Threshold" on page 438 descriptions.)
Key Path: Peak Search
State Saved: Not part of saved state.
Remote Command:
:CALCulate:MARKer[1]|2|3|4:MAXimum:LEFT
Example:
CALC:MARK2:MAX:LEFT selects marker 2 and moves it to the next peak to the left.