524 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
25 Trace/View
Notice that the displayed trace is now flat, or normalized.
3 The position of the normalized trace can now be moved to a different position on the
display by changing the normalized reference position.
This may be useful if the device to be tested has positive gain, such as an amplifier.
Now, replace the through line with the DUT and an accurate measurement of the gain or
loss can be made.
Key Path: Trace/View, Normalize
Dependencies/Couplings:
When Normalize is turned on, Volts, Watts, Amps, and Scale Type are not available.
If Scale Type is Linear, Normalize is not available.
State Saved: Saved in Instrument State.
Remote Command:
:CALCulate:NTData[:STATe] OFF|ON|0|1
:CALCulate:NTData[:STATe]?
Example:
CALC:NTD ON
CALC:NTD?
Norm Ref Lvl
Sets the level (in dB) of the normalized reference.
Key Path: Trace/View, Normalize
Factory Default: 0 dB
State Saved: Saved in Instrument State.
Range: 327.6 dB to +327.6 dB
Remote Command
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel <rel_ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel?
Example:
DISP:WIND:TRAC:Y:NRL .10 dB