System 24
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 477
Power On/Preset
Displays keys that enable you to define the instrument power-on state and user preset
state.
If Power On is set to Preset, and Preset Type is set to Factory, then turning on the N8201A
performs a factory preset. The last state of the N8201A (before it was turned off) is recalled
if Power On is set to Last. The user preset state is recalled if Power On is set to Preset and
Preset Type is set to User. (However, if there is no user preset state saved, the N8201A
performs a factory preset.)
Key Path: System
Power On
Enables you to set the state the N8201A will be put in when it is powered on. You can set
the state to Last or Preset.
If Power On is set to Last, then the instrument returns to the last saved instrument state
and all the modes are restored to that last state. The instrument saves its current state
internally every 2 minutes. So the last saved state, that is restored at power-on, is the state
at power-off or within a maximum of two minutes before power-off.
If Power On is set to Preset, the instrument state is determined by the preset type setting
prior to turning the power off.
Preset type Factory powers on in the Spectrum Analysis mode and all modes are set to
their factory defaults.
Preset type Mode powers on with the last mode the N8201A was in when it was
powered off and presets that mode to the factory defaults. It also restores all other
modes to their factory preset.
Preset type User powers on with the user defined state, saved when a Save User Preset
was last performed.
The setting (Last or Preset) of the Power On function is not changed by pressing Preset.
Use the Power On/Preset menu key function to change the setting of the N8201A state that
is recalled at power on. Limit lines are not recalled when the N8201A is powered on. Refer
to “Preset" on page 441 for more information.
Key Path: System, Power On/Preset
State Saved: Survives preset and power cycle, but not saved in Instrument State.
Remote Command:
:SYSTem:PON:TYPE PRESet|LAST
:SYSTem:PON:TYPE?