386 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Example:
SEM:BAND 4 MHz
SEM:BWID 4 MHz
SEM:BAND?
SEM:BWID?
SEM:BAND:AUTO 1
SEM:BWID:AUTO 1
SEM:BAND:AUTO?
SEM:BWID:AUTO?
Total Pwr Ref or PSD Ref
Displays the type of power measurement reference selected using the Meas Type key (see
“Meas Type" on page 383).
Total Pwr Re f is the power in the carrier that is used as the reference in computing the
relative power values for the offsets. When Man is selected, this can be set by the user.
When Auto is selected, this is the measured power in the reference carrier.
PSD Ref is the mean power spectral density in the carrier that is used as the reference in
computing the relative power spectral density values for the offsets. When Man is selected,
this can be set by the user, when Auto is selected, this is the measured mean power
spectral density in the reference carrier.
Key Path: Meas Setup, Ref Channel
Factory Preset: 0.00 dBm/Off
Range: –200.0 dBm to 200.0 dBm
Remote Command:
[:SENSe]:SEMask:CARRier[:POWer] <ampl>
[:SENSe]:SEMask:CARRier[:POWer]?
[:SENSe]:SEMask:CARRier:AUTO[:STATe] OFF|ON|0|1
[:SENSe]:SEMask:CARRier:AUTO[:STATe]?
Example:
SEM:CARR 0 dBm
SEM:CARR?