178 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
8 Frequency / Channel
Stop Freq
Sets the frequency at the right side of the graticule and sets the frequency entry mode to
Start/Stop. When the Start/Stop Frequency entry mode is activated, the start and stop
frequency values are displayed below the graticule in place of center frequency and span.
The left and right sides of the graticule correspond to the start and stop frequencies.
Key Path: FREQUENCY
State Saved: Saved in Instrument State
Factory Preset: 26.5 GHz
Range: –100 MHz to 27 GHz
Remote Command:
[:SENSe]:FREQuency:STOP <frequency>
[:SENSe]:FREQuency:STOP?
Example:
FREQ:STOP 1600
FREQ:STOP?
CF Step
Changes the step size for the center frequency and start/stop frequency functions. Once a
step size has been selected and the center frequency function is activated, the step keys
(and the UP|DOWN parameters for Center Frequency from remote commands) change
center frequency by the step-size value. The step size function is useful for finding
harmonics and sidebands beyond the current frequency span of the N8201A. When
auto-coupled in a non-zero span, the center frequency step size is set to 10% of the span.
Key Path: FREQUENCY
State Saved: Saved in Instrument State
Factory Preset: Span/10
Range: 1 Hz to 27 GHz
Remote Command:
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP[:INCRement]?