338 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Range Table
Enables you to enter the settings for up to 10 ranges, either using the instrument front
panel keys or remotely. Upon entering the range table, the measurement stops, then the
N8201A is set to a constantly sweeping state. The N8201A is then set to the current values
for range 1 (whether range 1 is on or off). If a range is currently off, the values in the range
table for that range are replaced with --- (see Range 5 in the illustration) to indicate this
range is currently inactive.
Key Path: Meas Setup
Saved State: All values for all ranges are saved in Instrument State.
Factory Preset: For each range, the following settings are selected: Range State(On),
Zero Span, Res BW Mode(Man), Video BW(Man) is set to 2 kHz, and Sweep Time
Mode(Auto) is set to 10 ms. The table below defines the other default parameter settings
for the measurement.
Table 6 Range Table Default Settings
Range Center
Freq (GHz)
Res BW
(kHz)
1 1.000 100.0
2 2.000 200.0
3 3.000 300.0
Comments to this Manuals