350 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Meas Method
Allows you to select the measurement method.
Above Threshold Lvl Selects the user defined threshold level or default level (–3.00 dB)
as the criteria in making the measurement.
Measured Burst Width This measurement method is not available for the following
radio standards: IS-95, J-STD-008, cdma2000-SR1, cdma2000-SR3, W-DCMA 3GPP.
The measurements described above are those available in SA mode (see Mode key). Other
measurements are available in other modes if an optional personality is installed.
Key Path: Meas Setup
Factory Preset: Above Threshold Lvl (THReshold)
Range: Measured Burst Width (BWIDth) to Above Threshold Lvl (THReshold)
Remote Command:
[:SENSe]:BPOWer:METHod THReshold|BWIDth
[:SENSe]:BPOWer:METHod?
Example:
BPOW:METH BWID to select burst width as the measurement method.