132 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
6 Display
Resolution: 0.1 dBm
Step Key Increment: 10 dBm
Range: 140 dBm to 100 dBm
Connected To Previous Pt
A current point may be connected to the previous point by pressing Yes. No limit testing
is performed between disconnected points. Pressing this key when the Connected field is
selected toggles the connected value of the current point and increments the Point
number to allow entry or editing of the Frequency of the next point. If an arrow key is
pressed without entering a value, the current Connected value of the point is selected. If
an arrow key is pressed, the Point number automatically increments to allow entry of the
Connected value of the next point. If it is a new point, the
Frequency may be entered for
the new point.
Key Path:
Display, Limits, Limit 1, Edit
Display, Limits, Limit 2, Edit
Factory Preset and *RST: Yes
Step Key Increment: Toggles to other selection
Delete Point
Deleting the current limit point can be achieved by pressing Delete Point. You will be
prompted with the message If you are sure, press key again to delete.
Pressing
Delete Point again will delete the limit point.
Key Path:
Display, Limits, Limit 1, Edit
Display, Limits, Limit 2, Edit
Delete Limit
Deletes the current limit LINE.
Key Path:
Display, Limits, Limit 1
Display, Limits, Limit 2
Factory Preset and *RST: Off