Amplitude / Y Scale 2
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 57
Example 2, set the following:
Scale Type (Lin)
Y Axis Units, Volts
Ref Level, 100 mV (10 mV/div)
This sets the top line to 100 mV and the bottom line to 0 V, so each vertical division
represents 10 mV. Thus, if a point on trace 1 is on the fifth graticule line from the top, it
represents 50 mV and will read out remotely as 50.
Example:
UNIT:POW dBmV
UNIT:POW?
dBm
Sets the amplitude units to dBm.
Key Path: AMPLITUDE / Y Scale, Y Axis Units
Remote Command:
See“Y Axis Units" on page 56.
Example:
UNIT:POW DBM
dBmV
Sets the amplitude units to dBmV.
Key Path: AMPLITUDE / Y Scale, Y Axis Units
Remote Command:
See“Y Axis Units" on page 56.
Example:
UNIT:POW DBMV
dBmA
Sets the amplitude units to dBmA.