394 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Range: 1 to 1000
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth:IMULti
<integer>
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth:IMULti?
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth
[:RESolution]:AUTO OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:BANDwidth|BWIDth
[:RESolution]:AUTO?
Remote Command Notes: Comma separated list. When offset F is off, this will be 5
values. When offset F is on, this will be 6 values. n = 1 for BTS and n = 2 for MS. Default is
BTS.
Example:
SEM:OFFS:LIST:BAND:IMUL 40 or SEM:OFFS:LIST:BWID:IMUL 40
SEM:OFFS:LIST:BAND:IMUL? or SEM:OFFS:LIST:BWID:IMUL?
SEM:OFFS:LIST:BAND:AUTO 0 or SEM:OFFS:LIST:BWID:AUTO 0
SEM:OFFS:LIST:BAND:AUTO? or SEM:OFFS:LIST:BWID:AUTO?
Abs Start
Enables you to enter an absolute level limit at Start Freq for the currently selected offset
ranging from -200.00 to +50.00 dBm with 0.01 dB resolution.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits" on page 388 for this value.
Default Terminator: dBm
Range: –200 dBm to 50 dBm
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:ABSolute
<ampl>,...[<ampl>] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:STARt:ABSolute?