Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 349
Threshold Lvl
Enables you to set the level above which the mean carrier power calculation is based. The
threshold level can be described in dB (Rel) or dBm (Abs).
Key Path: Meas Setup
Factory Preset: -30 dB
Te rm in a tor s: dB or dBm
Default Terminator dB
Range:
–60 dBm to 60 dBm (in absolute mode)
–60 dB to 0 dB (in relative mode)
Remote Command:
[:SENSe]:BPOWer:THReshold <number>
[:SENSe]:BPOWer:THReshold?
[:SENSe]:BPOWer:THReshold:TYPE ABSolute|RELative
[:SENSe]:BPOWer:THReshold:TYPE?
Example:
SENS:BPOW:THR:TYPE ABS