248 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
14 Measure (Spectrum Analysis Mode)
Multi-Carrier Power
Multi-Carrier Power is the measure of the power that leaks into adjacent transmit channels
when two or more carriers are present. The results reported are identical to the adjacent
power measurement, but the setup is different to allow for two or more carriers. For more
information see “Meas Setup (Multi-Carrier Power—MCP)" on page 307.
The measurement settings may be configured by pressing Meas Setup after Multi-Carrier
Power has been selected. Pressing Meas Control allows you to pause or restart your
measurement, or toggle between continuous and single measurement mode.
Key Path: MEASURE
Remote Command:
Measurement Results Available
Command n Results Returned
:CONFigure:MCPower N/A Not Applicable
:FETCh:MCPower[n]?
:MEASure:MCPower[n]?
:READ:MCPower[n]?
n=1 (or not
specified)
Returns 20 comma-separated scalar results, in the following order.
1 0.0
2 Total carrier power (dBm)
3 0.0
4 Reference carrier power (dBm)
5 lower 5 MHz offset - relative power (dBc)
6 lower 5 MHz offset - absolute power (dBm)
7 upper 5 MHz offset - relative power (dBc)
8 upper 5 MHz offset - absolute power (dBm)
9 lower 10 MHz offset - relative power (dBc)
10 lower 10 MHz offset - absolute power (dBm)
11 upper 10 MHz offset - relative power (dBc)
12 upper 10 MHz offset - absolute power (dBm)
13 lower 15 MHz offset - relative power (dBc)
14 lower 15 MHz offset - absolute power (dBm)
15 upper 15 MHz offset - relative power (dBc)
16 upper 15 MHz offset - absolute power (dBm)
17 NaN returned
18 NaN returned
19 NaN returned
20 NaN returned
If the results are not available, −999.0 is returned.
n=2...13 Returns absolute and relative values for carrier n-1. If results are not
available −999.0 is returned