86 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
3 Auto Couple
ADC Dither
Auto Turns the ADC dither to automatic. It then chooses on or off according to which is
most likely to be the best selection, based on the other N8201A settings such as span,
resolution BW and sweep type (FFT or swept).
On When ADC Dither is On, the linearity of low-level signals is improved. However, the
ADC dynamic range is reduced to make room for the dither. As a result, the noise floor of
the N8201A is somewhat compromised. So making measurements with ADC dither on
gives you better amplitude linearity, but turning ADC dither off gives you a lower noise floor
(better sensitivity).
Off When ADC Dither is Off, the instrument noise floor is improved. If ADC dither is on the
ADC dynamic range is reduced to make room for the dither. As a result, the noise floor of the
N8201A is somewhat compromised. So making measurements with ADC dither on gives you
improved amplitude linearity, but turning ADC dither off gives you a lower noise floor (better
sensitivity).