500 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
24 System
:SYSTem:LKEY? <“option”>
Remote Command Notes:
The query returns a string that contains the license key for a specified application or option
that is already installed in the instrument. The license key will also be returned if the
application is not currently in memory, but had been installed at some previous time. The
license key is unique to a specific option, host ID and serial number. Host ID can be returned
by :SYSTem:HID?.
Example:
SYST:LKEY “B78”, “B62A35B37679”
SYST:LKEY? “B78”
The query would return “B62A35B37679”
If the instrument does not have a license key for that option, the query would return ““.
Activate License
Activates the specified option. When a valid option key number and license key number
have been entered, press Activate. If your entry was correct, the option will be enabled and
the message: Option activated will appear in the status line of your display. The option
number and license key number will also be cleared from the License Key and Option key.
If your entry was incorrect, the error message: License key invalid will appear in the status
line of your display. The option number and license key number can be edited until they are
correct.
It is possible to enable an option for which the N8201A is not yet configured. Therefore, the
message Option activated does not mean that the option will immediately function. For a
display of options that are enabled and for which the N8201A is properly configured, press
System, More, Show System.
Key Path: System, Licensing
Remote Command:
There is no remote command for this key.
Delete License
Deletes the license key from memory, however, the option firmware is not deleted. The
instrument must be rebooted before the option is removed.