Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 343
Auto Fill Range Table
Automatically sets up the range table based on the current range center frequency and
span. When selected, the current center frequency is taken as the fundamental and the
frequency in range one is set to this value. Each of the subsequent range center frequency
parameters are set to the appropriate integer multiple of the fundamental frequency. The
span for range one is set to that of the current range. Each of the subsequent range span
parameters are set to the appropriate integer multiple of the fundamental span. This means
that if the fundamental range is zero span, all the ranges will be zero span. If the range span
is zero, Res BW, Video BW, and Sweep Time are set to the current base instrument
settings. If the range span is non-zero, Res BW, Video BW, and Sweep Time for all
harmonics are set to the same settings as for the fundamental. Span is set to zero for all
harmonics and the fundamental.
Key Path: Meas Setup, Range Table
Range: Sweep time range of your N8201A.
Remote Command: [:SENSe]:HARMonics:RTABle:FILL <no query>
Example: HARM:RTAB:FILL