180 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
8 Frequency / Channel
Freq Offset
Enables you to input a frequency offset value to account for frequency conversions external
to the N8201A. This value is added to the display readout of the marker frequency, center
frequency, start frequency, stop frequency and all other absolute frequency settings in the
N8201A. When a frequency offset is entered, the value appears below the center of the
graticule. Offsets may be entered using the keyboard. To eliminate an offset, perform a
Factory Preset or set the frequency offset to 0 Hz.
This command does not affect any bandwidths or the settings of relative frequency
parameters such as delta markers or span. It does not affect the current hardware settings
of the N8201A, but only the displayed frequency values. Offsets are not added to the
frequency count readouts. Entering an offset does not affect the trace display.
Key Path: FREQUENCY
State Saved: Saved in Instrument State
Factory Preset: 0 Hz
Range: –500 THz to 500 THz
Remote Command:
[:SENSe]:FREQuency:OFFSet <freq>
[:SENSe]:FREQuency:OFFSet?
Example:
FREQ:OFFS 10 MHz
Page view 179