260 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
14 Measure (Spectrum Analysis Mode)
Spurious Emissions
The spurious emissions measurement identifies and determines the power level of
spurious emissions in certain frequency bands. For more information see “Meas Setup
(Spurious Emissions)" on page 359.
Key Path: MEASURE
Remote Command:
Example:
MEAS:SPUR?
Command n Results Returned
:CONFigure:SPURious
:INITiate:SPURious
n/a Not applicable
:FETCh:SPURious[n]?
:MEASure:SPURious[n]?
:READ:SPURious[n]?
n=1 (or not
specified)
Returns a variable-length list of values containing detailed spur
information. The total number of data values returned is (1 + 6x
spurs), where x is the number of spurs identified. You can have up
to 121 data entries if you have the maximum 20 spurs identified.
The values returned are in the following order:
Number of spurs in items 2 through 6 in this list
[Repeat the following for each spur]
Spur number
Range number in which spur was located (integer)
Frequency of spur (Hz)
Amplitude of spur (dBm)
Absolute limit (dBm)
Pass or Fail (1|0)
n=2 through
21
Returns Detector 1 trace data values for the selected frequency
range (where range number = n 1). If the selected range is not
active, then 9.91E37 is returned for each non-active trace data
element.
n=22 Returns the number of spurs found.
n=23 through
42
Returns Detector 2 trace data values for the selected frequency
range (where range number = n 22). If selected range is not
active, then 9.91E37 is returned for each trace data element.