98 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
4 BW/Avg
Average
Initiates a digital averaging routine that averages the trace points in a number of successive
sweeps, resulting in trace “smoothing.” You can select the number of sweeps (average
number) with the keyboard. Increasing the average number further smooths the trace. To
select the type of averaging used, press BW/Avg, Avg/VBW Type.
Averaging restarts when any of the following occurs:
a new average number is entered.
any measurement related parameter (for example, center frequency) is changed.
Restart is pressed.
Single Sweep is pressed.
In single sweep, the specified number of averages is taken, then the sweep stops.
In continuous sweep, the specified number of averages is taken, then the
averaging continues, with each new sweep averaged in with a weight of
and the old average reduced by multiplying it by
.
To turn off averaging, press Average (Off).
Key Path: BW/Avg
Saved State: Saved in Instrument State
Factory Preset: Off, 100 averages
Range: Count: 1 to 8192
Remote Commands:
[:SENSe]:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:AVERage[:STATe]?
[:SENSe]:AVERage:COUNt <integer>
[:SENSe]:AVERage:COUNt?
[:SENSe]:AVERage:CLEar
Remote Command Notes:
For valid average data, you must restart the trace at the beginning of a sweep. To do this
remotely, first abort (:ABORT) the sweep and then initiate a single sweep (:INIT:CONT
OFF).