Agilent Technologies N8201A Service Manual Page 81

  • Download
  • Add to my manuals
  • Print
  • Page
    / 622
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 80
Auto Couple 3
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 81
PhNoise Opt
Selects the LO (local oscillator) phase noise behavior for various operating conditions. The
currently selected value is displayed below the £(f) indicator on the left side of the screen. It
is preceded by # if Auto has been selected (#£(f)). The key labels indicate measurement
situations where a specific type of optimization is best.
Key Path: Auto Couple
State Saved: Saved in Instrument State
Factory Preset: Auto
Remote Command: [:SENSe]:FREQuency:SYNThesis 1|2|3
1, selects optimization of phase noise for measuring signals with frequency offset
<50 kHz from the carrier.
2, selects optimization of phase noise for measuring signals with frequency offset
>50 kHz from the carrier.
3, selects optimization of LO phase noise for fast tuning (that is, faster
measurements)
[:SENSe]:FREQuency:SYNThesis?
[:SENSe]:FREQuency:SYNThesis:AUTO OFF|ON|0|1
[:SENSe]:FREQuency:SYNThesis:AUTO?
Example:
FREQ:SYNT:AUTO OFF
FREQ:SYNT 3, selects optimization for fast tuning
Auto
Selects the LO phase noise behavior to optimize speed or dynamic range for various
instrument operating conditions.
The Auto rules choose:
Fast Tuning, for span Š 10.5 MHz or the Res BW > 200 kHz
Optimize £(f) for f >50 kHz, for spans >141.4 kHz, and for Res BWs >9.1 kHz
Optimize £(f) for f <50 kHz, for all other spans and Res BWs.
Key Path: Auto Couple, PhNoise Opt
Page view 80