328 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
15 Measurement Setup
Meas BW
Sets the measurement 3 dB bandwidth.
Key Path: Meas Setup
State Saved: Saved in Instrument State.
Factory Preset: Defined by the radio standard selected.
Te rm in a tor s: Hz, kHz, MHz, GHz
Range: 10 kHz to 5MHz
Remote Command:
[:SENSe]:PSTatistic:BANDwidth|BWIDth <freq>
[:SENSe]:PSTatistic:BANDwidth|BWIDth?
Remote Command Notes:
The entered value is rounded to the nearest Res BW filter frequency.
This command is coupled with the N8201A resolution bandwidth.
Example:
PST:BAND 10 or PST:BWID 10