Agilent Technologies N8201A Service Manual Page 138

  • Download
  • Add to my manuals
  • Print
  • Page
    / 622
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 137
138 Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz
6 Display
Active Fctn Position
Selects the screen position for the Active Function Display. Depending on the type of trace
date being viewed, you can move the Active Function Display position for less visual
interference with your screen data.
Key Path: Display
Readback: The selection of the position to display the active function (Top, Center,
Bottom) reads back to line 3 of this key.
State Saved: Not saved in instrument state, survives power cycle and preset.
Factory Preset: No effect, persistent variable, survives power cycle and preset.
Factory Default: Center. Pressing System, Restore Sys Defaults will reset to factory
defaults.
Remote Command:
:DISPlay:AFUNction:POSition BOTTom|CENTer|TOP
:DISPlay:AFUNction:POSition?
Example:
:DISP:AFUN:POS BOT
Top
Displays the active function in the top-left corner of the display.
Page view 137
1 2 ... 133 134 135 136 137 138 139 140 141 142 143 ... 621 622

Comments to this Manuals

No comments