Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 283
OBW Span
Enables you to specify the range of integration used in calculating the total power from
which the percent occupied bandwidth is then calculated. The N8201A span will be set to
the same value as the OBW Span for the measurement. OBW Span should be set to
approximately 2 times the expected occupied bandwidth result.
If you have an adjacent channel, you should not set the OBW span to 2X your occupied
bandwidth. The OBW measurement first computes all the power in the span and then 99%
of that. Diamond markers are set around the bandwidth, and the occupied bandwidth
results is displayed in the data window. If the power of the adjacent channel is included in
the calculation for the 100% power, the OBW result will be too high. The OBW Span should
be set narrow enough to encompass the channel of interest and exclude any unwanted
adjacent channels.
Key Path: Meas Setup
Factory Preset: 3 MHz, or as defined by the radio standard selected.
Default Terminator: Hz
Remote Command:
[:SENSe]:OBW:FREQuency:SPAN <freq>
[:SENSe]:OBW:FREQuency:SPAN?
Example:
OBW:FREQ:SPAN 10 MHz
OBW:FREQ:SPAN?