Measurement Setup 15
Agilent N8201A Performance Downconverter Synthetic Instrument Module, 250 kHz to 26.5 GHz 395
Remote Command Notes: Comma separated list. When offset F is off, this will be 5
values. When offset F is on, this will be 6 values. n = 1 for BTS and n = 2 for MS. Default is
BTS.
Example:
SEM:OFFS:LIST:STAR:ABS –20 dBm
SEM:OFFS:LIST:STAR:ABS
Abs Stop
Enables you to enter an absolute level limit at Stop Freq for the currently selected offset
ranging from -200.00 to +50.00 dBm with 0.01 dB resolution, and to toggle this function
between Couple and Man. If set to Couple, this is coupled to Abs Start to make a flat limit
line. If set to Man, Abs Start and Abs Stop you can enter different values to make a sloped
limit line.
Key Path: Meas Setup, Offset/Limits
State Saved: Saved in Instrument State.
Factory Preset: Dependent upon Radio Standard and device selected, refer to the default
tables for “Offset/Limits" on page 388 for this value.
Default Terminator: dBm
Range: –200 dBm to 50 dBm
Remote Command:
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute
<ampl>,...[<ampl>] (up to five values)
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute?
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute:COUPle
OFF|ON|0|1
[:SENSe]:SEMask:OFFSet[n]:LIST:STOP:ABSolute:COUPle?
Remote Command Notes: Comma separated list. When offset F is off, this will be 5
values. When offset F is on, this will be 6 values. n = 1 for BTS and n = 2 for MS. Default is
BTS.
Example:
SEM:OFFS:LIST:STOP:ABS –20 dBm
SEM:OFFS:LIST:STOP:ABS