2-16 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
Measurement Modes
Multi Channel Pulsed Sweep Measurements
Multi channel pulsed sweep measurement is performed as shown below. The source
channels apply the pulsed sweep, staircase sweep, pulsed bias, or DC bias output,
and the measurement channels perform one point measurement at each sweep step.
Up to ten channels can be used for both pulsed sweep output and measurement. Both
voltage output mode and current output mode are available for the source channels.
Figure 2-10 Multi Channel Pulsed Sweep Measurements
1. Hold time, pulse period, measurement delay time, and number of measurements
are set by the MCPT command.
Sweep mode and number of sweep steps are set by the MCPWS command. They
are the common parameters for all sweep sources.
Trigger (e.g. XE)
MCPT, MCPWS, MCPNT, MCPWNX, WM
Peak value
Base value
Previous value
Time
: Measurement
Hold time
Delay time
Stop value
Start value
Previous value
Stop value
Base value
Previous value
Voltage or current
Start value
WNX
MCPNT, MCPNX
SMU1, N=1
SMU2, N=2
SMU9, N=9
Measurement delay time
Stop value
Base value
Previous value
Start value
MCPNT, MCPWNX
SMU8, N=8
Pulse period
Pulse width
Measurement time (set by AIT 2)
Stop value
Start value
Previous value
WNX
SMU3, N=3
DV
Setup value
Previous value
SMU10
Pulse period