Agilent B1500A/B1505A Programming Guide, Edition 11 1-39
Programming Basics
Data Output Format
Linear search,
Binary search
Data_search [Data_sense]<terminator>
This is the data at the measurement point closest to the search target. Data_search is
the value forced by the search output channel. Data_sense is the value measured by
the search monitor channel. It is sent if data output is enabled by the BSVM
command for the binary search, or the LSVM command for the linear search.
Staircase sweep,
Multi channel
sweep,
Multi channel
pulsed sweep,
CV (DC bias)
sweep
Block1 [Block2] . . . . <terminator>
Block1 is the block of data measured at the first sweep point. Block2 is the block of
data measured at the second sweep point. where Block consists of the following
data:
Data1 [Data2] . . . . [Source_data]
DataN (N: integer) is the value measured by a channel. The order of Data is defined
by the MM command. Source_data is the sweep source output value. It is sent if the
data output is enabled by the FMT command.
For the CMU measurement data, DataN will be Para1 Para2 [Osc_level Dc_bias].
Para1 and Para2 are R (Ω) and X (Ω), or G (S) and B (S) respectively. They will be
automatically selected by the B1500, and will be a couple without data overflow.
And Osc_level and Dc_bias are the monitor values of the oscillator level (AC signal
level) and the DC bias respectively. They are sent if the data output is enabled by the
LMN command.
Pulsed sweep,
Staircase sweep
with pulsed bias,
Pulsed sweep CV,
CV (AC level)
sweep,
C-f sweep
Block1 [Block2] . . . . <terminator>
Block1 is the block of data measured at the first sweep point. Block2 is the block of
data measured at the second sweep point. Block consists of the following data:
Data [Source_data]
Data is the value measured by the channel you specify by using the MM command.
Source_data is the sweep source output value. It is sent if the data output is enabled
by the FMT command.
For the pulsed sweep CV measurement, Data
will be Para1 Para2.
For the CV (AC level) and C-f, Data will be Para1 Para2 [Osc_level Dc_bias].
Para1 and Para2 are R (Ω) and X (Ω), or G (S) and B (S) respectively. They will be
automatically selected by the B1500, and will be a couple without data overflow.
And Osc_level and Dc_bias are the monitor values of the oscillator level (AC signal
level) and the DC bias respectively. They are sent if the data output is enabled by the
LMN command.