Agilent B1500A/B1505A Programming Guide, Edition 11 4-129
Command Reference
MCPNX
For HR/MP/HPSMU, available delay time value is 0. Also, the pulse width value
must be the same. If a different value is entered, the longest value is set.
MCPNX
The MCPNX command specifies the pulsed bias source and its parameters. This
command is effective for the multi channel pulsed spot or sweep measurement set
by MM 27 or MM 28.
To set the timing of output pulse and measurement, use the MCPT, MCPNT, and
AIT commands.
Syntax MCPNX N,chnum,mode,range,base,pulse[,comp]
Parameters N : Source number. Integer expression. 1 to 10. See Remarks below.
chnum : SMU pulsed source channel number. Integer expression. 1 to 10 or 101
to 1001. See Table 4-1 on page 4-16.
mode : Source type. Integer expression. 1 or 2.
1: Voltage source
2: Current source
range : Ranging type. Integer expression.
For voltage source (mode=1): See Table 4-4 on page 4-21.
For current source (mode=2): See Table 4-5 on page 4-22.
The B1500 usually uses the minimum range that covers both base and
pulse values to perform pulse output. For the limited auto ranging, the
instrument never uses the range less than the specified range.
base,
pulse: Pulse base or peak value (in V or A). Numeric expression.
For voltage source (mode=1): See Table 4-7 on page 4-24, Table
4-9 on page 4-26, Table 4-10 on page 4-26, Table 4-12 on page
4-27, Table 4-14 on page 4-28, or Table 4-15 on page 4-28 for each
measurement resource type. For using 3000 V range of HVSMU,
base and peak must have the same polarity.
For using UHVU, base and peak must have the same polarity.