Agilent Technologies B1500A Service Manual Page 398

  • Download
  • Add to my manuals
  • Print
  • Page
    / 598
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 397
4-82 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
ERHPL
1 (enable) or 0 (disable). Integer expression.
Example
Statements
OUTPUT @B1500;"ERHPE?"
ENTER @B1500;A
ERHPL
The ERHPL command enables or disables the LED status indicator of Agilent
N1258A/N1259A module selector.
Execution
Conditions
Digital I/O port must be set to the N1258A/N1259A control mode using the
ERMOD 2 command.
Syntax ERHPL onoff
Parameters onoff : 1 (enable, initial setting) or 0 (disable, always off). Integer expression.
Example
Statements
OUTPUT @B1500;"ERHPL 0"
ERHPL?
The ERHPL? command returns the condition of the LED status indicator of Agilent
N1258A/N1259A module selector.
Syntax ERHPL?
Query Response onoff<CR/LF^EOI>
1 (enable) or 0 (disable, always off). Integer expression.
Example
Statements
OUTPUT @B1500;"ERHPL?"
ENTER @B1500;A
ERHPP
The ERHPP command sets the input-to-output connection path of Agilent
N1258A/N1259A module selector.
Execution
Conditions
Digital I/O port must be set to the N1258A/N1259A control mode using the
ERMOD 2 command.
Syntax ERHPP path
Parameters path : Input-to-output connection path. Integer expression. 0, 1, 2, 3, or 4.
Page view 397
1 2 ... 393 394 395 396 397 398 399 400 401 402 403 ... 597 598

Comments to this Manuals

No comments