4-180 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
SPT
1: Pulse output active or ALWG sequence active
Example
Statements
OUTPUT @B1500;"SPST?"
ENTER @B1500;A
SPT
This command sets the pulse timing parameter for the specified SPGU channel. Set
for each channel. For the parameters, see Figure 2-31 on page 2-57.
Execution
Conditions
The SPGU operating mode must be set to PG with the SIM 0 command.
Syntax SPT chnum,src,delay,width,leading[,trailing]
Parameters chnum : SPGU channel number. Integer expression. 1 to 10 or 101 to 1002. See
Table 4-1.
src : Channel signal source. Integer expression.
1: Pulse signal source 1
2: Pulse signal source 2
delay : Pulse delay time. Numeric expression. 0 to pulse period-2E-8
seconds, setting resolution 2.5E-9 seconds. Initial setting 0 second.
Setting resolution becomes 1E-8 when leading > 8E-6 seconds or
trailing > 8E-6 seconds.
width : Pulse width. Numeric expression.1E-8 to pulse period-1E-8 seconds,
setting resolution 2.5E-9 seconds. Initial setting 1E-7 seconds.
Setting resolution becomes 1E-8 when leading > 8E-6 seconds or
trailing > 8E-6 seconds.
leading : Pulse leading time. Numeric expression. 8E-9 to 0.4 seconds, setting
resolution 2E-9 seconds. Initial setting 2E-8 seconds.
Setting resolution becomes 8E-9 when leading > 8E-6 seconds or
trailing > 8E-6 seconds.
trailing : Pulse trailing time. Numeric expression. 8E-
9 to 0.4 seconds, setting
resolution 2E-9 seconds. Initial setting 2E-8 seconds.
Setting resolution becomes 8E-9 when leading > 8E-6 seconds or
trailing > 8E-6 seconds.