Agilent B1500A/B1505A Programming Guide, Edition 11 2-41
Remote Mode Functions
Measurement Modes
NOTE The WTACV command sets the hold time, delay time, and step delay time.
The WMACV command sets the automatic abort function and the output after
measurement.
The WACV command sets the AC signal level sweep output.
The FC command sets the AC signal frequency.
The DCV command applies the specified DC bias.
If the SCUU (SMU CMU Unify Unit) is connected to the CMU and two
MP/HRSMU modules correctly, the source module is automatically selected by the
DC bias setting. The CMU is selected if it is ±25 V or less (setting resolution: 0.001
V), or the SMU is selected if it is greater than ±25 V (setting resolution: 0.005 V).
To Stop Sweep
Output
An automatic abort function is available. Refer to “Automatic Abort Function” on
page 2-46.