Agilent Technologies B1500A Service Manual Page 363

  • Download
  • Add to my manuals
  • Print
  • Page
    / 598
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 362
Agilent B1500A/B1505A Programming Guide, Edition 11 4-47
Command Reference
BGI
The compliance polarity is automatically set to the same polarity as the
stop value, regardless of the specified Icomp value. If stop=0, the
polarity is positive.
Remarks The time forcing the stop value will be approximately 1.5 ms to 1.8 ms with the
following settings:
BDM, BDT command parameters: interval=0, mode=0, delay=0
AV or AAD/AIT command parameters: initial setting
Example
Statements
OUTPUT @B1500;"BDV 1,0,0,100,0.01"
BGI
The BGI command sets the current monitor channel for the binary search
measurement (MM15). This command setting clears, and is cleared by, the BGV
command setting.
This command ignores the RI command setting.
Syntax BGI chnum,mode,condition,range,target
Parameters chnum : SMU search monitor channel number. Integer expression. 1 to 10 or
101 to 1001. See Table 4-1 on page 4-16.
mode,
condition : Search mode (0: limit mode or 1: repeat mode) and search stop
condition. The meaning of condition depends on the mode setting.
range : Measurement ranging type. Integer expression. The measurement range
will be set to the minimum range that covers the target value. For the
limited auto ranging, the instrument never uses the range less than the
specified range. See Table 4-3 on page 4-19.
mode condition
0 Limit value for the search target (target). The search stops when the
monitor data reaches target ± condition. Numeric expression. Positive
value. in A. Setting resolution: range/20000. where range means the
measurement range actually used for the measurement.
1 Repeat count. The search stops when the repeat count of the operation
that changes the source output value is over the specified value.
Numeric expression. 1 to 16.
Page view 362
1 2 ... 358 359 360 361 362 363 364 365 366 367 368 ... 597 598

Comments to this Manuals