Agilent Technologies B1500A Service Manual Page 44

  • Download
  • Add to my manuals
  • Print
  • Page
    / 598
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 43
1-26 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Basics
Data Output Format
ASCII Data Output Format
This section describes the ASCII data output format, and the elements of the data.
•“Time Stamp
•“Data Format
“Data Elements”
Time Stamp
The B1500 can record the time when the measurement is started, and sends the time
data (Time). This function is enabled by the TSC command. The time data will be
sent just before the measurement data.
For example, in the staircase sweep measurements, the data will be as shown below.
Block1 [,Block2] . . . . <terminator>
where, BlockN (N: integer) = Time1,Data1 [,Time2,Data2] ... [,Source_data], then
TimeN (N: integer) is the time from timer reset to the start of DataN measurement.
The timer count is cleared/reset by the TSR command (Time=0).
The time stamp function is not available for the following measurements.
High speed spot measurement
Quasi-pulsed spot measurement (MM9)
Linear search measurement (MM14)
Binary search measurement (MM15)
Page view 43
1 2 ... 39 40 41 42 43 44 45 46 47 48 49 ... 597 598

Comments to this Manuals

No comments