Agilent B1500A/B1505A Programming Guide, Edition 11 4-161
Command Reference
QSO
2: Stop value.
If this parameter is not set, the sweep sources force the start value.
Output Data The B1500 returns the data measured before an abort condition is detected. Dummy
data 199.999E+99 will be returned for the data after abort.
Example
Statements
OUTPUT @B1500;"QSM 2"
OUTPUT @B1500;"QSM 2,2"
QSO
This command enables or disables the QSCV smart operation for the quasi-static
CV measurement (MM13).
Execution
Conditions
The QSCV measurement operation must be Normal (QSC 0).
Syntax QSO mode[,chnum[,Vcomp]]
Parameters mode : QSCV smart operation. Integer expression. 0 or 1.
0: Disables the smart operation. Initial setting.
1: Enables the smart operation.
The QSCV smart operation is effective for the quasi-static CV
measurement with large leakage current, and needs an additional SMU.
The SMU must be specified by chnum.
Do not specify the following parameters when mode=0.
chnum : SMU source channel number. Integer expression. 1 to 10 or 101 to
1001. See Table 4-1 on page 4-16.
The specified SMU is used to perform the QSCV smart operation. The
SMU performs the current force operation to minimize the
measurement error caused by an offset current.
Vcomp : Voltage compliance (in V). Numeric expression. See Table 4-6 on page
4-23. If you do not specify Vcomp, the previous value is set.
This value is set to the SMU specified by chnum.
Example
Statements
OUTPUT @B1500;"QSO 1,5,20"