3-18 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
Staircase Sweep Measurements
Staircase Sweep Measurements
To perform staircase sweep measurements, use the following commands.
NOTE If you use multiple measurement channels, the channels start measurement in the
order defined in the MM command.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Sets integration time
(Agilent B1500 can use
AAD/AIT instead of AV.)
[AV] number[,mode]
[AAD] chnum[,type]
[AIT] type,mode[,N]
Sets sweep source timing
parameter
[WT] hold,delay
[,sdelay[,tdelay[,mdelay]]]
Sets auto abort function [WM] abort[,post]
Sets voltage sweep source WV chnum,mode,range,start,stop,step
[,comp[,Pcomp]]
Sets current sweep source WI
Sets synchronous sweep
source
a
a. The WSV/WSI command must be entered after the WV/WI command.
[WSV] chnum,range,start,stop
[,comp[,Pcomp]]
[WSI]
Forces constant voltage DV, TDV chnum,range,output
[,comp[,polarity[,crange]]]
Forces constant current DI, TDI
Sets voltage measurement
range
[RV] chnum,range
Sets current measurement
range
[RI] chnum,range
[RM] chnum,mode[,rate]
Sets measurement mode MM 2,chnum[,chnum ... [,chnum] ... ]
Sets SMU operation mode [CMM] chnum,mode
Executes measurement XE