3-42 Agilent B1500A/B1505A Programming Guide, Edition 11
Programming Examples
Binary Search Measurements
Binary Search Measurements
To perform binary search measurements, use the following commands.
The BSV and BSI commands clear the previous source settings.
Send the BSI command before sending the BSSI command.
Send the BSV command before sending the BSSV command.
The BSI/BSSV commands or BSV/BSSI commands cannot be used together.
Function Command Parameters
Enables channels CN [chnum ... [,chnum] ... ]
Disables channels CL [chnum ... [,chnum] ... ]
Sets filter ON/OFF [FL] mode[,chnum ... [,chnum] ... ]
Sets series resistor ON/OFF [SSR] chnum,mode
Sets integration time
(Agilent B1500 can use
AAD/AIT instead of AV.)
[AV] number[,mode]
[AAD] chnum[,type]
[AIT] type,mode[,N]
Sets measurement mode MM 15
Selects output data [BSVM] output_data
Sets timing parameters [BST] hold,delay
Sets source control mode BSM mode,abort[,post]
Sets current search or
voltage search condition
BGI or
BGV
chnum,mode,condition,range,
target
Sets voltage source or
current source
BSV or
BSI
chnum,range,start,stop[,comp]
Sets synchronous voltage
source or current source
[BSSV] or
[BSSI]
chnum,polarity,offset[,comp]
Forces constant voltage DV, TDV chnum,range
,output
[,comp[,polarity[,crange]]]
Forces constant current DI, TDI
Executes measurement XE
Page view 219