2-60 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
SPGU Module
NOTE Setup delay time and output voltage between the different patterns
When the pattern is changed to the pattern of the different index, 50 ns setup delay is
always inserted. There are no delay between the patterns of the same index.
During the pattern change, the channel keeps the last output voltage of the previous
pattern.
Figure 2-35 Trigger Output in ALWG Mode
NOTE SPUPD, ALS, and SRP commands
If the command is executed while the ALWG output continues, the SPGU will stop
the output and perform the channel setup operation again. After that, the SPGU will
start the output again if the operation mode is set to free run, and wait for the SRP
command if it is set to sequence count or duration.
SRP
SPP
Trigger Output
0 V (chassis common)
10 ns
STGP 901,1
SPGU 901
Pattern2
Pattern1
Pattern2
Pattern1
initial value
1 sequence
Pattern1
Trigger Output
0 V (chassis common)
10 ns
STGP 901,2
Trigger Output
0 V (chassis common)
>= 1 V
10 ns
STGP 901,3
>= 1 V
>= 1 V