Agilent B1500A/B1505A Programming Guide, Edition 11 4-107
Command Reference
*IDN?
Table 4-25 FMT mode parameter
*IDN?
The *IDN? query command returns the instrument model number and the firmware
revision number.
Syntax *IDN?
Query Response Agilent Technologies,model,0,revision<CR/LF^EOI>
Example
Statements
OUTPUT @B1500;"*IDN?"
ENTER @B1500;A$
Example
Response
Agilent Technologies,B1500A,0,A.04.03.2010.0130
IMP
The IMP command specifies the parameter measured by the MFCMU. This
command is effective for the following commands and the following measurement
mode. The MFCMU can measure two parameters.
mode Source data returned with measurement data
0 None (default setting). Only the measurement data is returned.
1 Source output data of the primary sweep source.
2 For MM2 and MM5:
Source output data of the synchronous sweep source set by the
WSI/WSV command.
1 to 10 For MM16, MM27, and MM28:
Source output data of the sweep source set by the WNX, MCPNX,
or MCPWNX command. The mode value must be the source
number (1 to 10) you want to get data. For the source number, see
description of the above commands.
Response Explanation
model B1500A or B1505A
revision Firmware revision number. Example: A.04.03.2010.0130