4-112 Agilent B1500A/B1505A Programming Guide, Edition 11
Command Reference
LIM?
Example
Statements
OUTPUT @B1500;"LIM 1,1500"
LIM?
This query command returns the voltage or current maximum output limit value of
SMU.
Syntax LIM? mode
Parameters mode : Type of the output limit value to read. 1 or 2. Integer expression.
1: Voltage limit value
2: Current limit value
Query Response limit<CR/LF^EOI>
This value returns the voltage output limit (V) or current output limit (A).
Example
Statements
OUTPUT @B1500;"LIM? 1"
ENTER @B1500;Limit
LMN
This command enables or disables the data monitor and data output of the MFCMU
AC voltage and DC voltage.
This command is effective for the spot C (MM17), DC bias sweep (MM18), and
frequency sweep (MM22), and AC level sweep (MM23).
Syntax LMN mode
Parameters mode : Data monitor and output mode. Integer expression. 0 or 1.
0: Disables the data monitor and output. Initial setting.
1: Enables the data monitor and output.
Example
Statements
OUTPUT @B1500;"LMN 1"
LOP?
The LOP? query command returns the operation status of all modules and stores the
results in the output data buffer (query buffer).