Agilent B1500A/B1505A Programming Guide, Edition 11 4-79
Command Reference
END
If unsupported error is detected, 999 is returned by the ERR? command.
The EMG? 999 command returns the message associated with the last error.
Syntax EMG? errcode
Parameters errcode : Error code returned by the ERR? command. Numeric expression.
Query Response error_message<CR/LF^EOI>
See Chapter 5, “Error Messages” for the error codes and error messages.
Example
Statements
OUTPUT @B1500;"EMG? 100"
ENTER @B1500;A$
See Also “ERR?”
END
The END command is used with the ST command to store a program in the internal
program memory. See “ST” on page 4-188.
Syntax END
Example
Statements
OUTPUT @B1500;"ST1;CN1;DV1,0,5,1E-4;TI1,0;CL1"
OUTPUT @B1500;"END"
ERC
The ERC command changes the output status of the digital I/O port. This command
does not change the status of the trigger ports and the input ports set by the ERM
command.
The *RST command or the device clear sets the digital I/O port (total 16 paths) to
the output port, and sets the port output level to TTL high.
Execution
Conditions
The digital I/O control mode must be the direct control (ERMOD 0).
Syntax ERC mode,value[,rule]
Parameters mode : Control mode. Integer expression. Set mode to 2.
2: Controls the digital I/O port.
If you set 1 that is effective for the Agilent 4142B, an error occurs.