RZ
range : Measurement range or ranging type. Integer expression. See Table 4-2
on page 4-17.
If you select the fixed range, the instrument performs measurement by using the
specified range.
For the auto or limited auto ranging, the measurement range will be set to the
minimum range that covers the measured values. However, the instrument never
uses the range less than the specified range for the limited auto ranging.
For the measurement mode that uses pulse source, if you select the auto or limited
auto ranging, the measurement channel uses the minimum range that covers the
compliance value or voltage output range.
Example
Statements
OUTPUT @B1500;"RV 2,-15"
OUTPUT @B1500;"RV 1,12"
RZ
The RZ command returns the channel to the settings that are stored by the DZ
command and clears the stored settings.
The DZ command stores the channel settings (V/I output values, V/I output ranges,
V/I compliance values, and so on), then sets the channel to 0 V.
Execution
Conditions
The channel setup has not been changed since the DZ command has been executed
for the specified channel. And the CL, CA, *TST?, *RST or a device clear (HP
BASIC CLEAR) command has not been executed for the specified channel.
Syntax RZ [chnum[,chnum...[,chnum]...]]
A maximum of 15 channels can be set.
Parameters chnum : Channel number. Integer expression. See Table 4-1 on page 4-16.
If you specify multiple chnums, the B1500 returns the stored settings in the order
specified.
If you do not specify chnum, this command returns the settings for all SMU, all
SPGU, and CMU that satisfy the conditions described in “Execution Conditions”
above, in this order. For the modules of same kind, the settings will be restored in
reverse order from their storage by the DZ command.
Example
Statements
OUTPUT @B1500;"RZ"
OUTPUT @B1500;"RZ 1,2,3"