Agilent B1500A/B1505A Programming Guide, Edition 11 4-71
Command Reference
DCV
corr : Correction mode. Integer expression. 1, 2, or 3.
1: Open correction
2: Short correction
3: Load correction
Query Response mode,primary,secondary<CR/LF^EOI>
mode : Measurement mode. Integer expression. 100 or 400.
100: Cp-G (for open correction)
400: Ls-Rs (for short or load correction)
primary : Primary reference value of the standard. Numeric expression.
Cp value for the open standard. in F.
Ls value for the short or load standard. in H.
secondary : Secondary reference value of the standard. Numeric expression.
G value for the open standard. in S.
Rs value for the short or load standard. in Ω.
Example
Statements
OUTPUT @B1500;"DCORR 9,3,400,0.00001,49.8765"
OUTPUT @B1500;"DCORR? 9,3"
ENTER @B1500;Mode,Primary,Secondary
This example returns Mode=400, Priamry=0.00001, Secondary=49.8765.
DCV
This command forces DC bias (voltage, up to ± 25 V) from the MFCMU. When the
SCUU (SMU CMU unify unit) is connected, output up to ± 100 V is available by
using the SMU that can be connected to the Force1/Sense1 terminals.
Execution
Conditions
The CN/CNX command has been executed for the specified channel.
If you want to apply DC voltage over ± 25 V, the SCUU must be connected
correctly. The SCUU can be used with the MFCMU and two SMUs (MPSMU or
HRSMU). The SCUU cannot be used if the HPSMU is connected to the SCUU or if
the number of SMUs connected to the SCUU is only one.
If the output voltage is greater than the allowable voltage for the interlock open
condition, the interlock circuit must be shorted.
Syntax DCV chnum,voltage