Agilent B1500A/B1505A Programming Guide, Edition 11 1-45
Programming Basics
Data Output Format
E Status. Three bits.
Status for Source_data:
Severity of a status is 001<010.
Status for measurement data. See Table 1-13 on page 1-46.
For SMU, the severity of a status is as follows:
For the quasi-pulsed spot measurement: 0<1<2<3<4<6 or 7
For other measurement: 0<6<7<1<2<3<4
F Channel number of the measurement/source channel. Five bits. See Table 1-14 on
page 1-47.
EExplanation
001 Data is for the first or intermediate sweep step.
010 Data is for the last sweep step.
Page view 62