2-58 Agilent B1500A/B1505A Programming Guide, Edition 11
Remote Mode Functions
SPGU Module
Figure 2-33 3-Level Pulse Output Examples
The SPGU output channels can be setup to be a 3-level pulse generator by using the
SPM 3 command. A 3-level pulse source can be made by defining two 2-level pulse
sources. For example, the upper example in Figure 2-33 can be made by the
following command sequence.
SPPER Period
SPM ch,3
SPT ch,1,Delay1,Width1,Leading1,Trailing1
SPV ch,1,0,1
SPT ch,2,Delay2,Width2,Leading2,Trailing2
SPV ch,2,0,-1
Pulseperiod
Width1
Width2
Delay2
Delay1
Output of
2-Level pulse source 2
Output of
2-Level pulse source 1
Pulseperiod
3-Level pulse output
Pulseperiod
Width1
Width2
Delay2
Delay1
Output of
2-Level pulse source 2
Output of
2-Level pulse source 1
Pulseperiod
3-Level pulse output
0 V
0 V
0 V
1 V
1 V
-1 V
-1 V
2 V
3 V
0 V
1 V
2 V
3 V
4 V